I. How to Operate the Equipment
(1) Please select the electrode type. (Touch Type, Adhesive Type).
(2) Please select the examinee’s posture type.
(Lying Posture, Seated Posture, Standing Posture)
(3) Please select whether you will be using ‘Dialysis Mode’ or not. (Enable, Disable)
If examinees are receiving kidney dialysis, make sure to set the equipment to ‘Dialysis Mode’.
Set the measurement time (before/during/after dialysis), access position, and paralysis location.
(4) Before the test, the examinee must maintain his/her posture for 10 minutes in order to reposition
his/her body water level.
(5) Please refer to Chapter D. Connecting the Electrodes for guidelines and attach the electrodes to the examinee’s
hands and feet.
(6) Input I.D., weight, height, age, and gender.
(7) Check that the electrodes are attached securely and press the ‘Enter’ button. If you did not attach the
electrodes to the examinee’s body, please do so now. After attaching the electrodes, please check again
to ensure that the examinee’s posture is correct and that the electrodes are connected securely. Once everything
has been properly set, press the ‘Enter’ button to start the test. Please make sure that the examinee maintains
his/her position during the test.
•
If you enter information that is out of range, the above error will appear
on the screen. Please re-enter the personal information.
Note