R E M O T E O P E R A T I O N S
Error Messages
36
LDC-3926339
C H A P T E R
3
E-211
Error in conversion to character pointer.
E-212
Error in conversion to byte pointer.
E-214
Response is too long to output.
E-222
Set value is over range.
E-223
Set value is under range.
E-226
Error in arbitrary block specification.
E-501
Interlock Open forces Laser output off.
E-503
Laser Voltage limit forces output off. "Open Circuit" error.
E-504
Laser Current Limit forces output off.
E-505
Laser Voltage Limit forces output off.
E-507
Laser Monitor Diode Power Limit forces output off.
E-508
External TEC Output Off Status forces LAS output off.
E-509
External TEC Temperature Limit forces LAS output off.
E-510
Out of Tolerance status forced LAS output off.
E-511
Laser Hardware Error forces output off. (e.g. low power line voltage).
E-529
Laser Output is off, but Laser status thought it was on.
E-535
Laser Mode changed while output on.
E-601
Internal error: recalled bin has incorrect checksum. (Settings do not match bin).
E-602
Internal error: task synchronization error.
E-620
Internal error: resource unavailable.
E-621
Internal error: message undeliverable to task.
E-622
Internal error: could not send message to mainframe.
E-710
AC Power Low Error detected
E-711
AC Power Low Error detected
E-713
Power Brown-Out Error detected
E-802
Calibration error: measurement entered before calibration was ready.
Table 3.4
Error Message Codes
Error Code
Explanation
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com