O P E R A T I O N S
Operating from the Front Panel
02_03
LDC-3926339
21
C H A P T E R
2
Note:
In this case, you may want to set the power limit (Plim) value high or set CalPD to
zero which disables the power limit feature.
7
Press the F1 softkey on any of the three pages of the LAS setup menu to turn on or off the
laser current source.
Also, you can turn on the current source with the F1 soft key from the CHAN menu. The
ON toggle has a two-second delay before the current source is enabled.
Turning On the Laser Current Source
1
Press the F1 softkey on any of the three pages of the LAS setup menu to toggle the laser
current source on and off.
Also, you can turn on the current source with the F1 soft key from the CHAN menu. The
ON toggle has a two-second delay before the current source is enabled.
2
Each laser current source can be turned on or off independently, or you can turn on all
channels of laser sources at the same time from the All Chnl menu. See the LDC-3926
mainframe manual for more information about all-channel operation.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com