A
XC1 Tubes
B
XC2 Tubes
C
TEM Tubes
D
STM Tubes
E
ATM Tubes
F
XC3 in Full Reservoir
G
RA1 in Half Reservoir
H
95% Formamide/1 mM EDTA in Quarter Reservoir
I
24 BeadChips in Chamber Rack
J
Temperature Probe
5
In the Basic Run Parameters pane, enter the number of BeadChips.
6
Select Run.
7
[Non-Illumina LIMS]
When prompted, enter the stain temperature listed on the STM tube. Do not load
the BeadChips yet.
8
When the chamber rack reaches 44°C, place the flow-through chambers into the chamber rack,
according to the robot deck map.
CAUTION
Start the robot immediately to prevent BeadChips from drying.
9
At the robot PC, select OK.
10 When the robot finishes, remove the flow-through chambers from the chamber rack, and place them
horizontally on the lab bench at room temperature.
Wash and Coat BeadChips
1
Gather the following equipment:
u
Kimwipes, large
u
Staining rack
u
Self-locking tweezers
u
Tube rack
u
Vacuum desiccator
u
Vacuum hose
u
Wash dishes (2)
2
During the procedure, prevent dust or lint from entering the wash dishes.
u
Clean wash dishes with low-pressure air before use.
u
Cover wash dishes with wash dish covers when not in use.
Document # 11322427 v03
For Research Use Only. Not for use in diagnostic procedures.
63
Infinium HD Super Assay Reference Guide