When the robot finishes, at the robot PC, select OK.
5
Remove the robot BeadChip alignment fixtures from the robot deck.
6
Make a record of any sections of BeadChip stripes without complete DNA sample coverage.
CAUTION
BeadChips must be removed from the robot deck immediately to prevent excess sample evaporation.
Proceed to the next section
(Set Up and Incubate BeadChips)
immediately.
Set up and Incubate BeadChips
1
Make sure that the Illumina Hybridization Oven is set to 48°C.
WARNING
Keep hybridization chambers at room temperature when you load the BeadChips. Do not place the
hybridization chamber in the Illumina Hybridization Oven when loading the BeadChips.
2
Open each hybridization chamber, and then carefully place each BeadChip in a hybridization chamber
insert.
Orient the barcode end so that it matches the barcode symbol on the Hyb Chamber Insert.
3
Make sure that hybridization chamber inserts are seated properly in the hybridization chambers.
4
Apply the back side of the hybridization chamber lid first, and then slowly bring down the front end to
avoid dislodging the hybridization chamber inserts.