4
Wait for the chamber rack to reach the correct temperature.
Stain BeadChips
1
If you are imaging the BeadChip immediately after the staining process, turn on the scanner to allow the
lasers to stabilize.
2
Fill the reservoir of each flow-through chamber as follows.
a
250 µl STM. Incubate for 10 minutes.
b
450 µl XC3. Incubate for 1 minute. Repeat 1 time..
c
Wait 5 minutes.
d
250 µl ATM. Incubate for 10 minutes.
e
450 µl XC3. Incubate for 1 minute. Repeat 1 time.
f
Wait 5 minutes.
g
250 µl STM. Incubate for 10 minutes.
h
450 µl XC3. Incubate for 1 minute. Repeat 1 time.
i
Wait 5 minutes.
j
250 µl ATM. Incubate for 10 minutes.
k
450 µl XC3. Incubate for 1 minute. Repeat 1 time.
l
Wait 5 minutes.
m 250 µl STM. Incubate for 10 minutes.
n
450 µl XC3. Incubate for 1 minute. Repeat 1 time.
o
Wait 5 minutes.
3
Immediately remove the flow-through chambers from the chamber rack and place in reserved alignment
fixtures submerged in PB1 at room temperature on a lab bench.
Wash and Coat BeadChips
1
Gather the following equipment:
u
Kimwipes, large
u
Staining rack
Document # 11322427 v03
For Research Use Only. Not for use in diagnostic procedures.
29
Infinium HD Super Assay Reference Guide