Mash Bath
I N D U S T R I A L E Q U I P M E N T & C O N T R O L P T Y . L T D .
6 1 - 6 5 M c C l u r e S t . T h o r n b u r y . 3 0 7 1 M e l b o u r n e . A u s t r a l i a
T e l : 6 1 ( 0 ) 3 9 4 9 7 2 5 5 5 F a x : 6 1 ( 0 ) 3 9 4 9 7 2 1 6 6
E m a i l : i e c @ i e c p l . c o m . a u
M A S H B - 2 0 1 9 0 3 1 4 . d o c 1 4 - M a r - 1 9
5
GENERAL INSTALLATION INFORMATION:
Bench mounting units are placed on the bench and are often used with sink to either the left
or right side of the machine. Floor machines are mounted on small wheels so that they can
easily be moved into and out of the laboratory location for cleaning or maintenance.
Electrical Services Required:
Standard voltages listed below. Machines frequency can be either 50 or 60 Hz. Please
contact IEC for other voltages if required.
BENCH MOUNTING MODELS:
SIZE
SINGLE
PHASE
AMP
THREE
PHASE
AMPS
/phase
HEATER
(bath)
SIZE WxDxH
cm
6 head
240V 1 ph
14
N/A
-
1x 3.1 kW
45x61x46
208V 1 ph
16
N/A
-
9 head
240V 1 ph
14
N/A
-
1x 3.1 kW
56x61x46
208V 1 ph
16
N/A
-
12 head
240V 1 ph
18
415V 3 ph
9
2x 2.0 kW
56x72x46
208V 1 ph
21
208V 3 ph
17
16 head
240V 1 ph
18
415V 3 ph
9
2x 2.0 kW
74x72x46
208V 1 ph
21
208V 3 ph
17
FLOOR MOUNTING MODELS:
SIZE
SINGLE
PHASE
AMP
THREE
PHASE
AMPS
/phase
HEATER
(bath)
HEATER
(d/w)
SIZE WxDxH
cm
16 head
240V 1 ph
29
415V 3 ph
12
2x 2.0 kW
2.8 kW
75x72x102
208V 1 ph
34
208V 3 ph
24
20 head
240V 1 ph
33
415V 3 ph
12
2x 2.5 kW
2.8 kW
75x83x102
208V 1 ph
39
208V 3 ph
24
25 head
240V 1 ph
36
415V 3 ph
12
2x 2.8 kW
2.8 kW
86x83x102
208V 1 ph
42
208V 3 ph
24
30 head
240V 1 ph
39
415V 3 ph
13
2x 3.1 kW
2.8 kW
97x83x102
208V 1 ph
45
208V 3 ph
27