Mash Bath
I N D U S T R I A L E Q U I P M E N T & C O N T R O L P T Y . L T D .
6 1 - 6 5 M c C l u r e S t . T h o r n b u r y . 3 0 7 1 M e l b o u r n e . A u s t r a l i a
T e l : 6 1 ( 0 ) 3 9 4 9 7 2 5 5 5 F a x : 6 1 ( 0 ) 3 9 4 9 7 2 1 6 6
E m a i l : i e c @ i e c p l . c o m . a u
M A S H B - 2 0 1 9 0 3 1 4 . d o c 1 4 - M a r - 1 9
2 3
This display permits editing, adding or
removal of any program name.
•
Press on the line you want to edit. A full
keyboard will appear with 3 alternative
screens to choose characters.
•
Print out the new name and press the
ENTER symbol to save the new name.
•
BACK returns to the Program Edit
display.
•
OFFSET/COOL enters another display
for editing.
EDIT PROG.NAME display
TEMPERATURE:: This display permits the
bath temperature to have a small OFFSET
applied to maximise accuracy. Use accurate
thermometer in the bath and adjust OFFSET
so the bath temp.control matches the
thermometer. Is normally within 0.5
o
C.
COOLING:: If bath temperature exceeds set
temperature by 2
o
C or more, the COOLING
function adds cold water to the bath. This
display permits adjustment of the cold water
time ON and time OFF for the best cooling
function. Typical: 5s ON, 5 to 10s OFF.
•
EXIT reverts to Program Editing.
•
STIRRING moves to editing stirring
functions.
OFFSET & COOLING display
STIRRING:: This display adjusts the speed
of all the magnetic stirrers. The speed for a
40mm long spin bar is set normally at about
400 RPM. This is about #14 for smaller
models and #16 for larger models. So all
stirrers run at the same speed, each bank of
stirrers under the bath has its own speed
adjustment control.
INTERMITTENT STIRRING: Some programs
require an intermittent stirring. This display
permits changing both the stir ON and the stir
OFF times.
Typical: 5s ON, 5 to 10s OFF.
•
EXIT reverts to Program Editing.
•
D/WATER moves to editing distilled
water functions (floor units only).
MAGNETIC STIRRING display