Mash Bath
I N D U S T R I A L E Q U I P M E N T & C O N T R O L P T Y . L T D .
6 1 - 6 5 M c C l u r e S t . T h o r n b u r y . 3 0 7 1 M e l b o u r n e . A u s t r a l i a
T e l : 6 1 ( 0 ) 3 9 4 9 7 2 5 5 5 F a x : 6 1 ( 0 ) 3 9 4 9 7 2 1 6 6
E m a i l : i e c @ i e c p l . c o m . a u
M A S H B - 2 0 1 9 0 3 1 4 . d o c 1 4 - M a r - 1 9
2 4
D/W TEMPERATURE:: For floor models,
this display permits the distilled water
temperature to have a small OFFSET applied
to maximise accuracy. Use accurate
thermometer in the d/w tank and adjust
OFFSET so the tank temp.control matches
the thermometer. Is normally within 0.5
o
C.
D/W DISPENSE ADJUSTMENT: The
dispense volume can be adjusted by either a
fixed ml ‘offset’ up or down (add or subtract ml
volume), OR by a percentage change in
volume up or down, OR both.
•
EXIT reverts to Program Editing.
•
P/WORD moves to editing the
password.
D/W TEMPERATURE display
ALTER PASSWORD:: When machine is
supplied, password is ‘1111’. This can be
altered to any other 4 digit password.
•
To alter any password at any time,
press the centre area. A keyboard will
appear asking for the current password to
be entered.
•
Enter 4 numeric digits and press the
ENTER symbol.
•
If valid, another display will appear
asking for the new password to be
entered.
•
EXIT reverts to Program Editing.
ALTER PASSWORD display
NEW PASSWORD:: When machine is
supplied, password is ‘1111’. This can be
altered to any other 4 digit password.
•
To enter the new password, press the
centre area. A keyboard will appear
asking for the NEW password to be
entered.
•
Enter 4 numeric digits and press the
ENTER symbol.
•
If valid, the display will revert to the
ALTER PASSWORD display.
•
EXIT reverts to Program Editing.
NEW PASSWORD display