9.2
Time Controller Commands
9.2.1
Device (DEVIce)
Common commands
Command tree
Input
Output
Description
DEVIce
:CALIbration:STATUS?
{SUCCESS
|FAILURE
|RUNNING }
Returns initial calibration status.
Restart the device in case of failure.
:CALIbration:EXECute
Recalibrates the Time Controller
:CONFIguration:LOAD
{COUNTercoinci
dence
|HISTOgram
|BLANK
|INIT}
-
Load predefined configurations:
-
COUNT: setup links as shown in
-
HISTO: restore links for histogram
and timestamps acquisition;
-
BLANK: setup all links to NONE
INIT: restore initial configuration
:LICEnse?
text
Returns the currently installed
license
:NETWork:ADDRess:MANUal
IPv4 address
(CIDR notation)
Set Time Controller manual (static)
IP address for MANUal mode.
:NETWork:ADDRess:MANUal?
IPv4 address
(CIDR notation)
Return the Time Controller manual
(static) IP address for MANUal mode.
:NETWork:ADDRess:MODE
{AUTO
|MANUal}
Set IP address allocation mode.
-
AUTO: from DHCP or APIPA
-
MANU: use manual IP address
:NETWork:ADDRess:MODE?
{AUTO
|MANUal}
Get network IP address allocation
mode.
:NETWork:RESEt
Apply network interface changes.
:LEDS
{ON|OFF}
Turns the LEDs interface ON or OFF
:LEDS?
{ON|OFF}
Returns the LEDs interface status
:RESolution
{HIRES
|LOWRES}
Select the device resolution mode
:RESolution?
{HIRES|LOWRES}
Returns the device resolution mode
:RESEt
Reset the Time Controller.
:SYNC
{INTernal
|EXTernal}
Set the device clock
location.
:SYNC?
{INTernal
|EXTernal}
:STATe?
State of the block