System Overview
3.1
Product Specification
The Time Controller consists of a single physical unit that can be connected to experimental
apparatus via five input channels and four output channels, while being controlled by a connected
host computer.
3.1.1
Functional Specification
3.1.1.1
Front Panel
Figure 1: Time Controller front panel
•
Power button:
This button switches the device on, switches the LEDs on and off with short
presses, and shuts the device down with presses longer than 2 seconds.
•
Start:
This is a signal input channel that is used when a start reference is required. Use an SMA
cable to connect a start signal output from a device to the start input in the Time Controller.
•
Inputs 1-4:
These are a signal input channels. Use an SMA cable to connect a signal output
from a device to an input channel on the Time Controller.
If you have selected the ‘Delay
Generator’ license for
your Time Controller, these input channels may be inaccessible (see
Section 3.1.6
–
•
Outputs 1-4:
These are signal output channels. Use an SMA cable to connect a signal output
channel to an input channel on the device. Each output can be used to generate a signal for
an input channel on the device.
If you have selected the ‘TCSPC’ license for your Time
Controller, these output channels may be inaccessible (see Section 3.1.6
–
•
LED Indicators:
Each input or output connector has an LED that indicates the status of the
input signal (see Section 5.2
Power
Start
(Input 0)
Input 1-4
Output 1-4