Time Controller Series User Manual
Section 6: Operating the Time Controller Software
ID QUANTIQUE SA
1227 Geneva
T +41 22 301 83 71
40
Rue Eugène-Marziano 25
Switzerland
F +41 22 301 83 79
Figure 12: Delay / pattern generator workspace
6.1.4
SCPI Console
The SCPI Console workspace can be used to manually issue SCPI commands to the Time Controller
device, load previously saved device configurations, and view the SCPI commands that the Time
Controller GUI has been pushing to the device.
Figure 13: SCPI Console workspace
To issue SCPI commands to the Time Controller device, type the command in the input field at the
bottom of the workspace and click
Send
.
To load a previously saved device configuration, click the folder icon on the right-hand side, and
select the desired configuration file. Device configuration files are generated by
navigating to ‘File >
Save Settings’
, saving the current device settings as a
‘
.bin
’
file with the name and location of your
choice.
Commands issued through the SCPI Console, or any other means outside of the
Time Controller GUI, will not be reflected in the Time Controller GUI. The
Time Controller software only registers a change in device parameter when the
command is issued via the Time Controller GUI.