4.2
Hardware Installation
Carry out the following steps to install the physical Time Controller unit, ready to power on and
operate with the host computer.
Warning: Electrostatic Shock Hazard
. Ensure all power connections are connected
to a ground socket. Failure to connect to ground the device creates a shock hazard
which may cause injury to the operator.
Warning: Trip Hazard
. Ensure all cables attached to the device are laud in a tidy
manner, fastened down if necessary. An untidied cable may represent a trip
hazard and cause injury to any users walking in the vicinity of the device.
Step 1
Place the Time Controller unit on a stable flat surface, in an environment meeting the
conditions specified in Section 3.1.3
–
.
Caution
: For optimal operation, ensure that the air inlet and air outlets are free of
any obstruction by at least 5 cm.
Step 2
Insert the supplied power cable into the power inlet on the rear panel of the device.
Step 3
Insert the other end of the supplied power cable into a grounded wall socket or
grounded uninterruptible power supply (UPS).
Step 4
Insert the ethernet cable into the 1GB Ethernet port on the rear panel of the device.
Step 5
Connect the other end of the ethernet cable to the host computer.
Note
: ID Quantique recommends use of the supplied ethernet-to-USB adapter,
even if an ethernet port is available on the host computer. Use of the adapter will
facilitate a smoother experience in operating the Time Controller software.
Step 6
Lightly screw each SMA connector onto the relevant Input, Output or Start channel on
the front panel of the device.
Caution
: Do not forcibly attach an SMA connector onto an SMA port. This may
permanently damage the SMA cable in question.
The device can now be powered on safely (see Section 5.1
–
Powering the Time Controller on and
.