Tip 1:
Error Code 18 Potential Causes
Dump valve activated, but water
level sensor does not read low
level probe within 5 minutes.
Water:
•
Is the dump valve faulty or in need of cleaning?
•
Does the dump valve have power from the PCB?
•
Is the drain line clogged with scale/debris?
•
Is the water level sensor faulty/bad?
•
Is the water pump faulty or in need of cleaning?
•
Is the water pump getting power from the PCB?
IM-Series
Error Codes
47