Program CAN channels
Observe the information in chapter
In the following example we show an assignment of a MultiXtend A with three
sensors:
•
Pressure sensor 0 – 600 bar at input 1
•
Pressure sensor 0 – 200 bar at input 2
•
Temperature sensor 0 – 60 °C at input 3
Program three special channels on the
MultiSystem 4070
, for example, chan-
nels 13 to 15.
In the
Calculation
field, select
MultiXtend
for each channel.
Activate MultiXtend power supply
MultiXtend
can either be supplied by its own power pack, or by the MS 4070
(CAN1 jack). If the instrument shall supply the required power, this function
must be activated:
1
Open the
Home
menu:
2
Open the
Setting
menu:
3
Open the
Device
menu:
4
Open the
Connections
menu:
5
Open the
CAN #1
dialog:
6
Select
Power supply
:
7
Set to
ON
:
8
Confirm changes and exit the dialog: