Firmware Version 2.0
MultiSystem 4070
113
Operating software
ENG
Setup Recording
In the
Setup Recording
dialog, you can predefine settings for the recording.
Recording start menu
Select whether the
Start recording
dialog is displayed before a recording
starts or whether the recording should start directly.
Mode
Choose from three options:
•
STANDARD
The defined recording and parameters will be applied to execute one sin-
gle recording
•
CYCLIC
The defined recording parameters will be applied to execute a recording;
then the recording will be repeated until the key
C-STOP
is pressed
•
SINGLE VAL
The current value of each selected channel will be recorded when key is
pressed
Data logger mode
Enable or disable data logger mode.
When data logger mode is enabled, the measuring instrument starts the recor-
ding directly after it is switched on.
For example, you can switch the instrument via the CAN jack so that it is star-
ted when you switch a machine on and recording begins.
OK
Confirms input/saves changes.