Firmware Version 2.0
MultiSystem 4070
112
Operating software
ENG
Example of a trigger
recording
A 2-minute recording is to be started when the measured value for p2 falls be-
low 50 bar and temperature T1 rises above 30 °C. The recording is to start 60
seconds before the trigger incident.
Required programming:
Recording time
2 min.
Trigger 1
p2
Trigger condi-
tion
FALLING
Trigger value
50.00
Pretrigger
50%
Trigger link
AND
Trigger 2
T1
Trigger condi-
tion
RISING
Trigger value
30.00