Firmware Version 2.0
MultiSystem 4070
109
Operating software
ENG
Trigger function
You can use the trigger function to reduce the amount of stored data by all-
owing the instrument to start the storing when the “interesting moments” are
coming. Here you can define up to four triggers.
Triggers are defined events that can start or stop a storing.
You may define any measurement channel as trigger, e.g. “if measured value
at channel 1 is greater than 10”, use a timer function, or use a manual key
press.
You can link four triggers logically, e.g. “if measured value at channel 1 is grea-
ter than 10 OR measured value at channel 2 is less than 100”. The trigger will
be started by the first of the two events.
Pretrigger
When a pretrigger is defined, the storing starts before the trigger event has
happened. The percentage defined as pretrigger is used to store measured
values before the trigger event.
Select a percent value as pretrigger –
.
OK
Confirms input/saves changes.