Performing inspection / maintenance
8. Carefully press the new filter element
into the element mount by applying slight
pressure, pressing it upwards and
screwing it in .
Do not use excessive force or a
hammer, etc.
9. (1) Before screwing the union nut back
on, apply lubricant to it.
We recommend using white Vaseline as
a lubricant
HYDAC part no. 632391
(2) Screw the filter bowl over the
coupling nut onto the filter head in a
counterclockwise direction.
When doing this, screw the union nut
hand tight only.
10. Put a membrane into the filter membrane
holder.
11. Put the test liquid that you previously
drained from the filter bowl into the
analysis chamber.
12. Start up the unit and clean the inner
chamber using the automatic cleaning
cycle with a volume of 5 ... 10 liters.
13. Remove the filter membrane from the
filter membrane holder.
14. Check the changed filters for leakage.
CTU 1000 Series
en(us)
Page 67 / 92
BeWa CTU1000 3229578p en-us 2017-04-25.doc
2017-04-25