Performing inspection / maintenance
Checking the gloves
The gloves are subject to wear both mechanically and chemically.
Mechanical wear - cuts and tears result from components with sharp edges
and wear can be accelerated by components with rough surfaces.
NOTICE
Gloves in the ultrasonic bath
The gloves will be destroyed
►
When the ultrasonic bath is in use, remove the gloves.
►
Pull the gloves through the openings to the outside.
For details of wear due to chemical action, for example by permeation,
please see page 76.
Provided there is no mechanical damage to the surface, the gloves can be
used for six months (with average use and single shift operation).
The tests that we carried out were based on the "G 60 Spezial" analysis
liquid.
The service life depends on the nature of the analysis liquid used and can
vary.
CAUTION
Worn gloves
Danger from test liquids, which can be
hazardous to your health
►
Check the gloves for leaks on a daily basis
►
If you find signs of wear, replace the gloves
immediately
►
Replace the gloves every six months