Error Messages and Troubleshooting
Error Messages and Troubleshooting
Error message
Cause(n)
Remedy
Invalid parameter
You have entered a value
which is outside the
permitted value range
(e.g. NAS 23).
Display the permitted
value ranges using the
keys
+
and
by
scrolling above the
maximum values. Then
the minimum value is
automatically displayed.
Check parameter
Defective
parameter:
No. xx
The self-monitoring
function of the FCU has
detected a check sum
error for one or more
parameters.
Cause: Strong
electromagnetic
interference has changed
the parameters.
Rest all parameters
using the keyboard or
transfer the parameters
with the FluMoS
software.
Measurement
ended
Memory for
measured values
full
The setting for the
"Memory mode"
parameter is at "Stop if
full".
The maximum number of
reports or measured
values has been reached.
No more values can be
stored.
Delete the reports or set
the "Memory mode"
parameter to "Overwrite".
For details see page .
Value not accepted
An error occurred while
storing a value in
EEPROM.
Repeat the entry or
continue the
measurement.
If the fault recurs,
contact HYDAC.