[HNT-1/1P] Data issue_Solenoid Control
[Adjust the Solenoid Control parameter]
1. It is recommended that the 'solenoid setup' is confirmed before the IOP calibration
2. Perform the pressure setup using the IOP 30 mmHg Model eye(27 ~ 32 mmHg)
3. Check the Applanation signal
4. In case of the “
NG
”, increase the Cur2 value to 165 and check again
Less than 2 lines
NG