[Execute the calibration for the Pachy measurement]
※ Please verify that the CUR REF value is 1.51 before execute the Pachy calibration
1. Set the model eye in the Model eye holder JIG
2. As shown first image, adjust the HNT-1P position so that the two line signal is visible
3. Press the “Red box” as shown below
4. As shown below second image, adjust the model eye position using the adjustment knob
on the holder JIG until the center of the model eye looks brighter overall
5. Press “Red box” for return to previous screen
[HNT-1/1P] Data issue_Pachy Calibration
5) How to do calibration for Pachy