Field Application Department
2018 . 10 . 02 (r1)
28/63
[Execute the calibration for the IOP measurement]
1. Press the table space to start IOP setup
- 10 ~ 30 mmHg Model eye setup
- Set up
in order
2. Set the target IOP of the current model eye
- Adjust target IOP with “+”, “-” buttons
1
3
2
4) How to do calibration for IOP
[HNT-1/1P] Data issue_IOP Calibration
1
3
2