General
771402-2 dopplex ABIlity Service Manual
6
2.4
Symbols
WARNING: .....
Indicates the possibility of death or serious injury
CAUTION: ......
Indicates the possibility of personal injury or material damage.
Attention, consult accompanying documents / Instructions for Use
2.5
Warnings
WARNING: A possible explosion hazard exists if used in the presence of flammable
anaesthetics.
WARNING: Do not operate the unit from the mains supply if the mains cable is damaged.
WARNING: Do not immerse any portion of the unit in water or other liquids.
WARNING: If this product is connected to another item of electrical equipment, it is
important that the system is fully compliant with EN60601-1-1.
WARNING: When configuring the system, consider and minimise the risk of persons tripping
over the tubing and electrical cables
2.6
Cautions
Observe ESD (electrostatic discharge) precautions when working within the unit and/or when
disassembling and reassembling the dopplex Ability and when handling any of the
components of the dopplex Ability.
2.7
System Connection
WARNING: These requirements must be met when a dopplex Ability is connected to any
other electrical equipment, such as a PC.
Non-medical equipment must comply with the relevant IEC or ISO safety standard. For Information
Technology equipment, this standard is IEC950/ EN60950.
Medical equipment must comply with IEC601-1/EN60601-1, or equivalent.
The configured system must comply with the requirements of IEC60601-1:2005; clause 16.
If non-medical equipment (e.g. the PC or printer) with enclosure leakage currents greater than those
allowed by IEC601-1/ EN60601-1 is to be used in the patient environment (within 1.5m of the patient),
the enclosure leakage currents must be brought within the limits laid down by IEC601-1/EN60601-1.
This may be achieved by using a medical grade isolating transformer. Suitable types are available via
Huntleigh sales agents.
Anyone who connects additional equipment to signal input or signal output parts of the system is
configuring a medical system, and is therefore responsible for ensuring that the system complies with
the requirements of IEC60601-1:2005; clause 16. If there is any doubt as to whether your system
complies, consult the technical service department of your local Huntleigh representative.
Summary of Contents for DA100PB
Page 1: ...771402 2 dopplex ABIlity Service Manual 1 SERVICE MANUAL ...
Page 15: ...Product Description 771402 2 dopplex ABIlity Service Manual 15 ...
Page 22: ...System Block Diagram 771402 2 dopplex ABIlity Service Manual 22 6 7 Pneumatic Circuit Diagram ...
Page 23: ...System Block Diagram 771402 2 dopplex ABIlity Service Manual 23 ...
Page 24: ...System Block Diagram 771402 2 dopplex ABIlity Service Manual 24 ...
Page 29: ...Product Specifications 771402 2 dopplex ABIlity Service Manual 29 8 4 General ...
Page 66: ...Fault Modes 771402 2 dopplex ABIlity Service Manual 66 Figure 3 Inflation Envelope ...