Disassembly & Assembly
771402-2 dopplex ABIlity Service Manual
35
9.5
KEYPAD (SP-771335) REPLACEMENT
Remove the foam tape and unplug the keypad
flexi-cable.
Carefully remove the faulty keypad from the Display
panel.
Locate the new keypad over the Display panel.
Push the flexi-cable (FFC) tail through the recess in the
Display panel
Remove the backing from the keypad.
When in position ensure all surfaces are in contact
by pressing all edges firmly.
Plug in the keypad flexi-cable and replace the foam tape.