Disassembly & Assembly
771402-2 dopplex ABIlity Service Manual
46
9.16 Over Pressure PCB (SP-771507) Replacement
Desolder the 8 pins of the Over-pressure PCB.
Using solder braid clean out the 8 PCB holes.
Put the Overpressure Detect PCB in place
with the unpopulated side towards the pump.
Solder in place using lead free solder
.
Summary of Contents for DA100PB
Page 1: ...771402 2 dopplex ABIlity Service Manual 1 SERVICE MANUAL ...
Page 15: ...Product Description 771402 2 dopplex ABIlity Service Manual 15 ...
Page 22: ...System Block Diagram 771402 2 dopplex ABIlity Service Manual 22 6 7 Pneumatic Circuit Diagram ...
Page 23: ...System Block Diagram 771402 2 dopplex ABIlity Service Manual 23 ...
Page 24: ...System Block Diagram 771402 2 dopplex ABIlity Service Manual 24 ...
Page 29: ...Product Specifications 771402 2 dopplex ABIlity Service Manual 29 8 4 General ...
Page 66: ...Fault Modes 771402 2 dopplex ABIlity Service Manual 66 Figure 3 Inflation Envelope ...