Disassembly & Assembly
771402-2 dopplex ABIlity Service Manual
34
9.4
Base PCB (SP-7711067) Removal
Remove the 2 off M3x16 countersunk screws.
On reassembly torque to 90cNm
Remove the 2 off KB30-10 pan head screws
On reassembly torque to 60cNm
Unplug the Battery cable from the PCB.
Unplug the tubing from the pump to the base.
Slide the Base PCB forwards so that it clears the 2
plastic locating tabs then lift off the Base PCB.
Summary of Contents for DA100PB
Page 1: ...771402 2 dopplex ABIlity Service Manual 1 SERVICE MANUAL ...
Page 15: ...Product Description 771402 2 dopplex ABIlity Service Manual 15 ...
Page 22: ...System Block Diagram 771402 2 dopplex ABIlity Service Manual 22 6 7 Pneumatic Circuit Diagram ...
Page 23: ...System Block Diagram 771402 2 dopplex ABIlity Service Manual 23 ...
Page 24: ...System Block Diagram 771402 2 dopplex ABIlity Service Manual 24 ...
Page 29: ...Product Specifications 771402 2 dopplex ABIlity Service Manual 29 8 4 General ...
Page 66: ...Fault Modes 771402 2 dopplex ABIlity Service Manual 66 Figure 3 Inflation Envelope ...