Product Description
771402-2 dopplex ABIlity Service Manual
18
Patient Type Set-Up Menu Map
5.6
System Connection
WARNING: These requirements must be met when a dopplex Ability is connected to any
other electrical equipment, such as a PC.
1. Non-medical equipment must comply with the relevant IEC or ISO safety standard. For Information
Technology equipment, this standard is IEC950/ EN60950.
2. Medical equipment must comply with IEC601-1/EN60601-1, or equivalent.
3. The configured system must comply with the Medical System standard IEC601-1-1/ EN60601-1-1.
4. If non-medical equipment (e.g. the PC or printer) with enclosure leakage currents greater than those
allowed by IEC601-1/EN60601-1 is to be used in the patient environment (within 1.5m of the patient),
the enclosure leakage currents must be brought within the limits laid down by IEC601-1/EN60601-1.
This may be achieved by using a medical grade isolating transformer. Suitable types are available via
Huntleigh sales agents.
5.
Anyone who connects additional equipment to signal input or signal output parts of the system is
configuring a medical system, and is therefore responsible for ensuring that the system complies with
IEC60601-1-1/EN60601-1-1. If there is any doubt as to whether your system complies, consult the
Service Department of your local Huntleigh representative.
Summary of Contents for DA100PB
Page 1: ...771402 2 dopplex ABIlity Service Manual 1 SERVICE MANUAL ...
Page 15: ...Product Description 771402 2 dopplex ABIlity Service Manual 15 ...
Page 22: ...System Block Diagram 771402 2 dopplex ABIlity Service Manual 22 6 7 Pneumatic Circuit Diagram ...
Page 23: ...System Block Diagram 771402 2 dopplex ABIlity Service Manual 23 ...
Page 24: ...System Block Diagram 771402 2 dopplex ABIlity Service Manual 24 ...
Page 29: ...Product Specifications 771402 2 dopplex ABIlity Service Manual 29 8 4 General ...
Page 66: ...Fault Modes 771402 2 dopplex ABIlity Service Manual 66 Figure 3 Inflation Envelope ...