Safety Testing
771402-2 dopplex ABIlity Service Manual
52
10.3 Electrical Safety Test: USB (Enclosure Leakage) Test
10.3.1
Connect the unit under test as illustrated below to a suitable electrical safety analyser.
10.3.2
Switch on the unit.
10.3.3
Allow the unit to fully initialise before applying the tests in the table below.
Test
Test Limit
Normal Polarity
100
μA
No L2
500
μA
Reversed Polarity
100
μA
No L2
500
μA
Live
Earth
Neutral
Neutral from Analyser
Earth from Analyser
Live from Analyser
Enclosure Socket of Analyser
USB Socket