![background image](http://html1.mh-extra.com/html/human-care/55100/55100_user-manual_2182698031.webp)
www.humancaregroup.com
31
ENGLISH
Hanger Bars
Lifting height
Connect the hanger bar to the hook on the
lift belt by threading the hook diagonally from
beneath (see fig).
• 72754 Hanger bar HeliQ 2P 35 cm
• 72755 Hanger bar HeliQ 2P 45 cm
• 72757 Hanger bar HeliQ 2P 60 cm
• 92938 Hanger bar HeliQ 4P
The lifting height corresponds to the difference
between the highest and lowest position the lift
can reach. This corresponds to the length of the
lift belt. The length of the lift belt is 2120mm
(83,5”)/1720 mm (66.1”).
If you want to extend the lift belt there are
distance straps in various lengths available as
extras. The lifting height of the lift is not affec-
ted by the length of the distance strap.The total
height of the system consists of the rail in which
the lift is mounted, the rail roller in the rail, a
distance strap if relevant and the lift.
Accessories
If the maximum weight for any of the accessori-
es, e.g. slings, lift bars and any other accessories
is NOT the same as for the lift, then the lowest
max. weight given for any one product is the
one that shall apply. Always check the label on
the lift and lifting accessories or contact Human
Care if you have any questions or if anything is
unclear.
Slings
Human Care’s range of slings is wide enough
to meet the user’s various requirements and
needs. Slings are tested in accordance with Hu-
man Care’s test log, which can be downloaded
from www.humancaregroup.com. A test must
be performed for each client individually and in
accordance with their needs. Consult the user
manual for the relevant sling to make sure the
product is used correctly.
Human Care recommends using only Human
Care slings with Human Care lifts. Visit www.
humancaregroup.com for further information
about which models and sizes are offered.
Motorized operation
HeliQ offers the ability to connect to motorized
operation. Motorized operation allows the lift
to be moved along the rail with minimal effort
from the caregiver. Motorized operation needs
to be installed by a technician with Human Care
qualifications.
The power traverse is equipped with over-
heating protection which stops the motor if it
overheats. Overheating can occur if the motor
is running continuously during a longer period.
Continuous operation is maximum 5 minutes.
More than this will lead to overheating.