ENGLISH
Indicator
Charger
The indicator light has the following indica-
tions, which are only accessible when the lift is
switched on.
The indicator on the lift flashes red when the
batteries need recharging. Plug the charger
cable into the hand control or put the hand
control into the charger cradle and charging
will start. When the lift is charging the indicator
glows yellow.
The lift can only be used with a charger
recommended by Human Care, article
number 50880.
• Charge the lift regularly, preferably every
night
• The charger must be disconnected when
you use the lift
• The unit charges regardless of the position
of the main switch
• If the emergency stop is activated the lift
cannot be charged. Check that the emer-
gency stop is not activated when charging.
• The charger must only be connected to an
earthed socket.
If rail charging is available via the rail, the lift
will charge continuously when the lift is not in
use and is within the charging zone. When rail
charging is available, the lift does not need to
be charged via the hand control. Get in touch
with the system owner or Human Care if you
are uncertain as to what type of charging to
use.
Important! When the batteries
reach a critically low battery level,
it will no longer be possible to
perform more lifts (up or down).
However, there is still enough
battery power for emergency
lowering. See separate manual for
the charger, 99599.
Press the Up button on the hand control or
control panel on the lift to lift the client. Press
the Down button on the hand control or control
panel to lower the client. Pressing the button
activates a soft start, and releasing it causes a
soft stop.
Flashing
green
The lift is on and ready for use.
Battery level is Full.
Flashing
yellow
The lift is on and ready for use.
Battery level is Normal. It might be
a good idea to charge up the lift.
Flashing
red
The lift is on and ready for use.
Battery level is Low. The lift must be
charged up immediately.
Constant
green
The lift is in operation or loaded.
Constant red
The lift safety catch has engaged.
(See troubleshooting table).
Constant
yellow
The lift is charging.
Flashing
green/yellow
The lift’s service interval has passed.
Service and annual inspection must
be performed immediately.
Flashing red/
yellow
The lift’s service interval has passed.
Service and annual inspection must
be performed immediately. Battery
level is Low. The lift must be char
-
ged up immediately.