Guidance and manufacturer’s declaration
Guidance and manufacturer’s declaration – electromagnetic emissions
The Human Care lift is intended for use in the electromagnetic environment specified below. The customer or the
user of the Human Care lift should assure that it is used in such an environment.
Emission test
Compliance
Electro magnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Human Care lift uses RF energy only for its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The Human Care lift is suitable for use in all esta
-
blishments including domestic establishments and
those directly connected to the public low-voltage
power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions IEC
61000-3-2
Class A
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The Human Care lift is intended for use in the electromagnetic environment specified below. The customer or the
user of the Human Care lift should assure that it is used in such an environment.
Immunity test IEC 60601 test level
Compliance level
Electro magnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete or ceramic
tile.
If floors are covered with synthetic material,
the relative humidity should be at least 30 %.
Electrical fast
transient /
Burst
IEC 61000-4-4
+/- 2 kV for power
supply lines
+/- 1 kV for input/out
-
put lines
+/- 2 kV for power
supply lines
n/a. for input/output
lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
+/- 1 kV differential
mode
+/- 2 kV common
mode
+/- 1 kV differential
mode
n/a. for common
mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips,
short interrup
-
tions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip
in UT))
for 5 sec
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT))
for 5 sec
Mains power quality should be that of a typical
commercial or hospital environment. If the
user of the Human Care lift requires continued
operation during power mains interruptions, it
is recommended that the lift be powered from
an uninterruptible power supply or battery.