28
www.humancaregroup.com
HeliQ
Hand control
Secondary control panel
The hand control is equipped with an Up
button (2) and a Down button (3) for raising
and lowering the lift. The hand control is also
equipped with an OFF/ON button (1) for activa-
tion/deactivation of the lift and an emergency
lowering button (4) for quickly lowering the lift.
Push the emergency lowering button together
with the down button at least for 3 seconds
simultaneously to activate the emergency
lowering. Older lifts might not have the dual
control and the delay.
Lifts equipped with a Power Traverse have a
hand control with seven buttons. In addition
to the buttons described above, there are also
buttons for moving the lift forwards (7) and
backwards (6), as well as an alarm button (5).
When the lift is loaded with less than approx.
40 kg it can be run at two speeds. If the Up
or Down button is pressed for longer than 3
seconds, the lift will run in the faster mode. If
the lift is loaded with more than approx. 40 kg it
will only run in the slower mode.
On the short side of the hand control there is an
input for charging the lift and an input (USB-C)
for communicating with the lift.
The lift is equipped with a secondary control
panel located on the lift itself. This control panel
can be used as an alternative to the hand con-
trol. The secondary control panel is equipped
with an Up button (2) and a Down button (3)
for raising and lowering the lift. The control
panel is also equipped with an OFF/ON button
(1) for activation/deactivation of the lift and
an emergency lowering button (4) for quickly
lowering the lift.
Press emergency lowering button and down
button at least 3 seconds simultaneously to ac-
tivate the emergency lowering. Older lifters may
not have double key command and delay.
When the lift is loaded with less than approx.
40 kg it can be run at two speeds. If the Up
or Down button is pressed for longer than 3
seconds, the lift will run in the faster mode. If
the lift is loaded with more than approx. 40 kg
it will only run in the slower mode. The force
required to operate the push buttons is 4.4N.
1 2 3 4
HeliQ
HeliQ Power traverse
1
2
3
4
5
6
7