background image

2

II. Feature

High-voltage insulation resistance tester, with 7 range: 500V, 1000V, 2500V, 5000V,

10000V, 15000V, 20000V,test up to 20TΩ, manual voltage regulation function.
● Strict accordance with the safety standards for design

Insulation resistance range 20TΩ @ 20 kv

● Short circuit current can be adjusted, maximum 5mA (Can be ordered as high as 10mA,
short circuit current products).
● Automatic display polarization index (PI), induced electric absorption test values than
(DAR), and can test the leakage current and capacitance.
● Anti-jamming performance is superior, the interference current reaches 2 mA, still
ensure accuracy of test instruments.

2 methods for power: use of lithium ion batteries, battery life can be up to 4 hours

(10000V @ 100M

Ω

).

Insulation resistance simulation column.

Digital filtering functions, external influence deviation display value causing mitigation

point using filter function.

Perfect protection function, and the spare fuse prompt functions.

III. Specification Parameter

Safety Standard

IEC 61010-1

CAT.IV 600V

IEC 61326

EMC specification: test, control and electrical equipment for testing

IEC60529 IP64

Outer closed state

● AC Power

220V±10%

50/60 HZ

20 VA

● battery

16.8V Lithium ion rechargeable battery

● Battery Life

10000V@100M

4 hour

● size

L x W x H

):

37cm x 27cm x 17cm

● weight

3.6kg

● Test voltage accuracy: the nominal value 100% to 110%

● Voltage measurement accuracy

5%+3V

● Current test range

10mA

● current measure precision

5%+0.2nA

● short-circuit current

5mA

● Capacitance test range

20uF

● Capacitance testing accuracy

15%+0.03uF

Summary of Contents for HZJY-20K

Page 1: ...HZJY 20K Insulation Resistance Tester User Manual Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...l and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instr...

Page 3: ...Contents I Warning 1 II Feature 2 III Specification Parameter 2 IV Introduction of instrument 3 V Instrument using 6 VI Packing List 7...

Page 4: ...erators in instrument surface operation Please don t charged the connection test line low Low in the measurement or do not touch the circuit being measured immediately after the test may result in ele...

Page 5: ...ion batteries battery life can be up to 4 hours 10000V 100M Insulation resistance simulation column Digital filtering functions external influence deviation display value causing mitigation point usi...

Page 6: ...1T 2T range accuracy 15000V 20000V Unspecified 1M 1M 5 1M 300G 1M 400G 20 300G 3T 400G 4T Unspecified 3T 4T IV Introduction of instrument The Front Panel Description number Description number Descrip...

Page 7: ...case recommended 1 hardware low pass filter applicable to most situations 2 software Low pass filter High resistance greater than 100G recommended choose 3 hardware software low pass filter high resis...

Page 8: ...to determine the current Settings and exit Help menu system HELP Record the serial number of product production date version information Insulation testing interface Digital Insulation value greater t...

Page 9: ...opped display shows as shown in the table of the soft keys Soft key Description OFF Power Off No operation 3 minutes automatic shutdown Back Return to the main menu H V Set test voltage Manual 500V 10...

Page 10: ...een When measuring high insulation resistance the sample should be in the position a conductor surface protection between the two measuring end ring and the conductor protection ring is connected to t...

Reviews: