● Insulation resistance testing range and accuracy

T:23±5ºC

RH :45 – 75%

range

accuracy

500V

1000V

2500V

5000V

10000V

Unspecified

<1M

<1M

<1M

<1M

<1M

5%

1M -10G

1M -20G

1M -100G

1M -100G

1M -200G

20%

10G -100G

20G-200G

100G-1T

100G-1T

200G-2T

Unspecified

> 100G

> 200G

> 1T

> 1T

> 2T

range

accuracy

15000V

20000V

Unspecified

<1M

<1M

5%

1M -300G

1M -400G

20%

300G -3T

400G-4T

Unspecified

> 3T

> 4T

IV. Introduction of instrument

The Front Panel Description

number

Description

number

Description

1

AC plug

2

AC ON

3

Test Port

E

G

L

4

LCD display

5

Key

6

HV LED

Key and function

Key numbered from left to right, the first for a total of five off down the power Key.

key

number

Description

1

Press the switch on button startup, shutdown condition

Summary of Contents for HZJY-20K

Page 1: ...HZJY 20K Insulation Resistance Tester User Manual Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...l and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instr...

Page 3: ...Contents I Warning 1 II Feature 2 III Specification Parameter 2 IV Introduction of instrument 3 V Instrument using 6 VI Packing List 7...

Page 4: ...erators in instrument surface operation Please don t charged the connection test line low Low in the measurement or do not touch the circuit being measured immediately after the test may result in ele...

Page 5: ...ion batteries battery life can be up to 4 hours 10000V 100M Insulation resistance simulation column Digital filtering functions external influence deviation display value causing mitigation point usi...

Page 6: ...1T 2T range accuracy 15000V 20000V Unspecified 1M 1M 5 1M 300G 1M 400G 20 300G 3T 400G 4T Unspecified 3T 4T IV Introduction of instrument The Front Panel Description number Description number Descrip...

Page 7: ...case recommended 1 hardware low pass filter applicable to most situations 2 software Low pass filter High resistance greater than 100G recommended choose 3 hardware software low pass filter high resis...

Page 8: ...to determine the current Settings and exit Help menu system HELP Record the serial number of product production date version information Insulation testing interface Digital Insulation value greater t...

Page 9: ...opped display shows as shown in the table of the soft keys Soft key Description OFF Power Off No operation 3 minutes automatic shutdown Back Return to the main menu H V Set test voltage Manual 500V 10...

Page 10: ...een When measuring high insulation resistance the sample should be in the position a conductor surface protection between the two measuring end ring and the conductor protection ring is connected to t...

Reviews: