5

2. Press the 1 key (the power button), until into the contrast adjustment menu.
3. Press 1 or 2 keys to adjust contrast, until the right .
4. Press the 4 key save and exit, if you don't need to adjust the press 3 key to exit.

Soft key

Description

Move the cursor, also used to increase or decrease in contrast
Move the cursor, also used to increase or decrease in contrast

Back

Return to the main menu

OK

Confirm to enter the current menu function or to determine the
current Settings and exit

Help menu system

HELP

Record the serial number of product, production date, version information.

Insulation testing interface

Digital

Insulation value

( >

greater than

; <

less than

)

Measured time

actual voltage

H.V

warning

Leakage current

1mA=1000uA
1uA =1000nA

DAR

Capacitance

PI

Simulation column

When the test time or manual stopped, display shows as shown in the table of the soft

keys.

Soft key

Description

OFF

Power Off (No operation, 3 minutes automatic shutdown)

Back

Return to the main menu

H.V

Set test voltage:

Manual

500V

1000V

2500V

5000V

10000V

15000V

20000V

Rt

Veiw 15S

30S

60S

600S Insulation value

Summary of Contents for HZJY-20K

Page 1: ...HZJY 20K Insulation Resistance Tester User Manual Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...l and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instr...

Page 3: ...Contents I Warning 1 II Feature 2 III Specification Parameter 2 IV Introduction of instrument 3 V Instrument using 6 VI Packing List 7...

Page 4: ...erators in instrument surface operation Please don t charged the connection test line low Low in the measurement or do not touch the circuit being measured immediately after the test may result in ele...

Page 5: ...ion batteries battery life can be up to 4 hours 10000V 100M Insulation resistance simulation column Digital filtering functions external influence deviation display value causing mitigation point usi...

Page 6: ...1T 2T range accuracy 15000V 20000V Unspecified 1M 1M 5 1M 300G 1M 400G 20 300G 3T 400G 4T Unspecified 3T 4T IV Introduction of instrument The Front Panel Description number Description number Descrip...

Page 7: ...case recommended 1 hardware low pass filter applicable to most situations 2 software Low pass filter High resistance greater than 100G recommended choose 3 hardware software low pass filter high resis...

Page 8: ...to determine the current Settings and exit Help menu system HELP Record the serial number of product production date version information Insulation testing interface Digital Insulation value greater t...

Page 9: ...opped display shows as shown in the table of the soft keys Soft key Description OFF Power Off No operation 3 minutes automatic shutdown Back Return to the main menu H V Set test voltage Manual 500V 10...

Page 10: ...een When measuring high insulation resistance the sample should be in the position a conductor surface protection between the two measuring end ring and the conductor protection ring is connected to t...

Reviews: