Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 77 of 222
General Tab
Figure 8-3
Write the method’s name in the blank field on the top of the window (to the left of the Add button) then press
Add
. The method will appear in the field to the left of the
Delete
button, included in the general list of
methods.
NOTE: Do not change the size of the method name into the general list for one big. Software will
cut the method name if you change to another one too big.
For a new methods file, the name will appear in the field to the left of the
Add
button. For an already existing
methods file, you have to overwrite the name of any method in the field left of the
Add
button, enter the new
name and finally press
Add
Warning: When the first method is programmed, it will be displayed with the default values.
When a new method is created, be sure to enter all of the desired parameter values, since the
software will default to those values of the currently displayed method.
Fill in different fields in the “General” Tab:
Name
: Enter the generic method name; this name will appear in the printed reports exactly as it’s been
written in this field.
Brand
: Enter the name of the reagent manufacturer.
Type:
In this field, the methods are programmed according to the type of reaction.
Software has five types of reactions, and each has its own options and characteristics. Each one will be
explained in detail later.