Fibrin or salt is plugging the
electrode flow path.
Remove electrodes and clean or
replace electrode with plugged flow
path. Reinstall electrodes and
recalibrate.
Calibrant A pump is not working
properly.
Contact Technical Support.
(0733) ISE Module - Air in
Calibrant B (automatically
repeated)
Calibrant B is segmented with air
Electrodes are not properly seated.
Check compression plate, spring and
seal.
Ensure that all electrodes and o- rings
are properly installed.
Ensure tubing between reagent kit and
sample entry port is connected properly.
Replace tubing between reagent kit and
sample entry port.
Reagent low or out.
Fibrin or salt is plugging the
electrode flow path.
Use cleaning procedure.
Remove electrodes and clean or
replace electrode with plugged flow
path. Reinstall electrodes and
recalibrate.
Waste pump malfunction
Contact Technical Support.
Bubble detector malfunction
Contact Technical Support.
(0734) ISE Module
Temperature higher than
32ºC. Please decrease the
environment temperature,
otherwise the Electrodes life
expectancy could be
reduced
ISE Module Temperature higher
than 32ºC.
Room temperature must be reduced.
(0735) ISE Module
Temperature drift higher than
4ºC. Please Calibrate the
Module and repeat the test
Processing temperature is +/- 4ª
since last calibration.
Calibrate the ISE Module and repeat the
test.
(0751) ISE Module - Invalid
Answer
Internal error.
Contact Technical Support.
(0752) ISE Module -
Unknown Error
Internal error.
Contact Technical Support.
(0753) ISE Module - Air in
Sample
See possible solution for 0731.
(0754) ISE Module - Air in
Calibrant A
See possible solution for 0732.
(0755) ISE Module - Air in
Calibrant B
See possible solution for 0733.