(0764) ISE Module
calibration out of time.
Please Calibrate it
Calibration time expired (more
than 8 hours from last)
Run a calibration procedure.
(0765) ISE Module
uncalibrated. Please
Calibrate it
Module has been turned off.
Run a calibration procedure.
ISE module has been under Mant
command
Calibration ended with errors
Electrode has been enabled by
hardware option
(0766) ISE Module
communication Checksum
error
Internal error
Contact Technical Support.
(0767) ISE Module answer
parsing error
Internal error
Contact Technical Support.
(0768) ISE Module cleaning
out of time. Please Clean it
Cleaning time expired (more than
8 hours or 50 samples from last)
Run cleaning procedure.
(0769) ISE Module out of
position. Please reposition
the Module
ISE module in a wrong position.
Reposition module.
(0770) ISE Pack changed
ISE Reagent Kit has been
changed.
Run Reagent Kit replacement.
(0771) ISE Pack Security
Code error
Kit electronic key failure
Change ISE reagent kit.
(0772) ISE Pack Wrong
Distributor Code
Invalid Reagent Kit
5420-0104 ISE reagent kit must be
used.
(0773) ISE Pack Expired
ISE Reagent Kit expired.
Change ISE reagent kit. Run Reagent
Kit replacement
(0774) ISE Pack Not Good
Kit electronic key failure
Change ISE reagent kit. Run Reagent
Kit replacement
(0775) ISE Module in
Maintenance Mode. Run a
Purge Cycle or a Calibration
Cycle
ISE module in Maintenance
Mode.
Run Calibrant A Purge or Calibration
procedure.
(0776) ISE Module Off.
Position the module properly
and Reconnect it
ISE module is off.
Run Reconnect procedure and then
Star up.
(0777) ISE Pack not
connected
ISE Reagent Kit is not connected.
Connect ISE reagent kit
ISE Reagent Kit connector out of
place.
Check ISE reagent Kit connector.
(0778) ISE Pack no more
Calibrant
Lack of Calibrant
Replace ISE reagent kit.
(0779) ISE Module
Calibration successful
ISE Calibration finished
successfully
No action is required
(0780) Last ISE Module
Calibration finished with
warnings
Last ISE Calibration ended with
warnings
Check last ISE Calibration warning
messages, correct them and recalibrate
the ISE Module.