Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 115 of 222
In the lower left side, the quality control graph is shown. Programmed control values are shown in solid
line:
Gray = Mean value (this value is the average value between the programmed maximum and minimum
control values)
Green = ± 1SD (these values are obtained by considering the programmed maximum and minimum
control values as ± 2SD)
Yellow = ± 2SD (this values are obtained from the programmed maximum and minimum control values)
Red = ± 3SD (these values are obtained by considering the programmed maximum and minimum control
values as ± 2SD)
Selected controls statistics are shown in dotted line Gray = Mean value
Green = ± 1SD
Yellow = ± 2SD Red = ± 3SD
Figure 11-14
Lines can be individually hidden by pressing the related upper side reference. For example, by pressing on
the
-3SD Cont
text, the bottom red line will be hidden.
12 MAINTENANCE MENU
The
Maintenance Menu
is accessed from the main screen of software (Figure 12-1). In this menu general
configuration options of the system, maintenance operations and validation tests are separately described.