High Technology, Inc.
109 Production Road,
Walpole MA USA 02081
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared By
J.S. Bolio
Approved By
S. Titov
Effective date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
Page 126 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
-Washing 3: Aspiration of liquid and dispensing of washing solution
-Washing 4: Aspiration of liquid and dispensing of washing solution
-Washing 5: First drying step
-Washing 6: Second drying step
Light Blue: In Use Gray: Unknown status
Cuvettes status is constantly refreshed as operations are carried out in the cuvettes tray.
Warning: Every time the software is unexpectedly closed in unusual way, the instrument shows the
status of the cuvettes as all dirty, this is a preventive action.
Therefore, it is necessary to start sessions by washing dirty cuvettes or by carrying out the programmed
washing. Otherwise, start the new session by washing the entire cuvettes tray
(Wash cuvettes
range
of cuvettes
to 100
wash)
Figure 12-18