prog.book : ch7mem.fm 22 Mon Feb 15 09:24:37 1999
MEMory Subsystem
MEMory:TABLe:FREQuency <numeric_value>{,<numeric_value>}
7-22
HP E4418B/E4419B Programming Guide
MEMory:TABLe:FREQuency <numeric_value>{,<numeric_value>}
This command is used to enter frequency data into the current selected
table. Any previous frequency list is cleared before the new frequencies
are stored. The frequencies must be entered in ascending order. Entries in
the frequency lists correspond as shown with entries in the
calibration/offset factor lists. Note, that for sensor calibration tables only,
the first calibration factor entered using the
MEMory:TABLe:GAIN
command is used as the reference calibration factor.
For sensor calibration tables, the number of frequency points must be one
less than the number of calibration factor points. This is verified when the
sensor calibration table is selected using
SENSe:CORRection:CSET:SELect <string>
.
Ensure that the frequency points you use cover the frequency range of the
signals that you want to measure. If you measure a signal with a
frequency outside the frequency range defined in the table, then the power
meter uses the highest or lowest point in the table to calculate the
calibration factor/offset.
The power meter is capable of storing 20 sensor calibration tables and 10
frequency dependent offset tables, each containing 80 points.
Syntax
Frequency
Calibration Factor/Offset
-
Reference Calibration Factor
(For Sensor Calibration Tables )
Frequency 1
Calibration Factor/Offset 1
Frequency 2
Calibration Factor/Offset 2
"
"
Frequency 80
Calibration Factor/Offset 80
MEM
:TABL
:FREQ
numeric_value
Space
,
?
Summary of Contents for E4418B
Page 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Page 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Page 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Page 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Page 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Page 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Page 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Page 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Page 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Page 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Page 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Page 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Page 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Page 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Page 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Page 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...