prog.book : chapter1.fm 51 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Setting Offsets
HP E4418B/E4419B Programming Guide
1-51
Setting Offsets
Channel Offsets
The power meter can be configured to compensate for signal loss or gain in
your test setup (for example, to compensate for the loss of a 10 dB
attenuator). You use the
SENSe
command subsystem to configure the
power meter. Gain and loss correction are a coupled system. This means
that a gain set by
[SENSe[1]]|SENSe2:CORRection:GAIN2
is
represented in the
[SENSe[1]]|SENSe2:CORRection:LOSS2?
command. If you enter an offset value the state is automatically enabled.
However it can be enabled and disabled using either the
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe
or
[SENSe[1]]|SENSe2:CORRection:LOSS2:STATe
commands.
LOSS2
is coupled to
GAIN2
by the equation
when the default
unit is linear, and
when the default is logarithmic.
Note
You can only use
LOSS2
and
GAIN2
for external losses and gains.
LOSS1
and
GAIN1
are specifically for calibration factors.
Display Offsets
Display offset values can be entered using the
CALCulate[1|2]:GAIN[:MAGNitude]
command.
CALCulate[1|2]:GAIN:STATe
must be set to
ON
to enable the offset
value. If you enter an offset value the state is automatically enabled. On
the HP E4419B this offset is applied after any math calculations (refer to
Figure 1-9 on page 1-70).
Loss
1
Gain
-------------
=
Gain
Loss
–
=
Summary of Contents for E4418B
Page 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Page 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Page 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Page 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Page 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Page 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Page 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Page 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Page 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Page 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Page 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Page 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Page 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Page 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Page 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Page 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...