prog.book : chapter1.fm 37 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Using Sensor Calibration Tables
HP E4418B/E4419B Programming Guide
1-37
Reviewing Table Data
To review the data stored in a sensor calibration table, use the following
commands:
•
MEMory:TABLe:SELect "Sense1"
Select the sensor calibration table named “Sense1”.
•
MEMory:TABLe:SELect?
Query command which returns the name of the currently selected
table.
•
MEMory:TABLe:FREQuency:POINTs?
Query command which returns the number of stored frequency
points.
•
MEMory:TABLe:FREQuency?
Query command which returns the frequencies stored in the sensor
calibration table (in Hz).
•
MEMory:TABLe:GAIN[:MAGNitude]:POINTs?
Query command which returns the number of calibration factor
points stored in the sensor calibration table.
•
MEMory:TABLe:GAIN[:MAGNitude]?
Query command which returns the calibration factors stored in the
sensor calibration table. The first point returned is the reference
calibration factor.
Modifying Data
If you need to modify the frequency and calibration factor data stored in a
sensor calibration table you need to resend the complete data lists. There
are two ways to do this:
1.
If you have retained the original data in a program, edit the
program and resend the data.
2.
Use the query commands shown in “Reviewing Table Data”, on
page 1-37 to enter the data into your computer. Edit this data,
then resend it.
Summary of Contents for E4418B
Page 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Page 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Page 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Page 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Page 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Page 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Page 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Page 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Page 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Page 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Page 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Page 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Page 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Page 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Page 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Page 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...