Power Meter Remote Operation
Making Measurements
HP E4418B/E4419B Programming Guide
1-29
Using the Lower Level Commands
An alternative method of making measurements is to use the lower level
commands to set up the expected range and resolution. This can be done
using the following commands:
[SENSe[1]]|SENSe2:POWER:AC:RANGe
DISPlay[:WINDow[1|2]]:RESolution
The measurement type can be set using the following commands in the
CALCulate
subsystem:
CALCulate[1|2]:MATH[:EXPRession]
CALCulate[1|2]:RELative[:MAGNitude]
The advantage of using the lower level commands over the
CONFigure
command is that they give you more precise control of the power meter. As
shown in Table 1-3 on page 1-14 the
CONFigure
command presets various
states in the power meter. It may be likely that you do not want to preset
these states.
Example
The following example sets the expected power value to -50 dBm and the
resolution setting to 3 using the lower level commands. The measurement
is a single channel A measurement carried out on the lower window.
ABOR1
Aborts channel A.
CALC2:MATH:EXPR "(SENS1)"
Displays channel A on lower
window.
SENS1:POW:AC:RANG 0
Sets lower range (E-series sensors
only).
DISP:WIND2:RES 3
Sets the lower window’s resolution
to setting 3.
INIT1
Causes channel A to make a
measurement.
FETC2?
Retrieves the lower window’s
measurement.
Summary of Contents for E4418B
Page 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Page 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Page 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Page 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Page 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Page 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Page 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Page 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Page 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Page 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Page 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Page 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Page 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Page 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Page 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Page 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...