HP 81110/04A Quick Start
Getting Started
107
Getting Started
Examples
Device Command List for the Burst Signal
•
If you want to include this Burst Signal in your program use the fol-
lowing command lines in boldface font. The comment lines starting
with a # are not required.
NOTE:
A second channel is required.
# Reset the instrument to start from a defined, default status.
*RST
# Switch off the automatic display update to increase programming
# speed.
:DISPlay OFF
# Set the instrument to burst mode by selecting a burst count of 5.
# Choose double pulses for output 2. Select triggered mode by selecting
# PLL as the trigger source and set the burst repetition to 5
µ
s.
:TRIGger:COUNt 5
# Set Burst mode with burst count of 5
:ARM:SOURce INT2
# Set Triggered mode with PLL as
# trigger source
:ARM:PERiod 5US
# Set burst repetition of 5
µ
s
:PULSe:DOUBle2 ON
# second channel generates double
# pulses per period
# Set the period to 500 ns, for output 1 enter a width of 100 ns. For
# output 2 set the double pulse to 250 ns and the width to 100 ns. All
# edges set to 3 ns.
:PULSe:PERiod 500NS
Summary of Contents for 81104A