HP 81110/04A User’s Guide Installation & Safety Requirements
23
Installation & Safety Requirements
Ventilation Requirements
Ventilation Requirements
The instrument is fitted with three cooling fans. Make sure that there is
adequate clearance of 80 mm at the rear and 25 mm at the top and bot-
tom to ensure adequate air flow. If the air flow is restricted the internal
operating temperature will be higher, reducing the instrument’s reliabil-
ity or causing the instrument’s thermal-protection circuits to automati-
cally switch off the instrument.
NOTE:
Do not cover the ventilation holes.
Thermal Protection
Overheating Detection
The instrument monitors its internal temperature. If the temperature
exceeds approximately 80
°
C the power supply is switched off. The
instrument will switch on again if the temperature falls below approxi-
mately 77
°
C.
Fan Failure
If a fan is broken or prevented from operating by a blockage the temper-
ature will increase. When the temperature exceeds approximately 80
°
C
the overheating detection switches off the instrument for safety reasons.
For reliability it is recommended to send instruments with broken or
defective fans immediately to HP Service for repair.
Summary of Contents for 81104A
Page 1: ...HP 81110A 165 330 MHz HP 81104A 80 MHz Pulse Pattern Generators Quick Start Guide ...
Page 2: ...Front Panel Display and Softkeys ...
Page 8: ...10 About this book HP 81110 04A Quick Start About this book ...
Page 31: ...HP 81110 04A Quick Start Introduction 33 Chapter 2 2Introduction ...
Page 67: ...HP 81110 04A Quick Start Getting Started 69 Chapter 3 3Getting Started ...
Page 114: ...116 Index HP 81110 04 Quick Start Index ...
Page 115: ...Front Panel Controls ...