HP 81110/04A Quick Start
Getting Started
101
Getting Started
Examples
Device Command List for the Dual Clock Signal
•
If you want to include this Dual Clock Signal in your program use
the following command lines in boldface font. The comment lines
starting with a # are not required.
NOTE:
A second channel is required.
# Reset the instrument to start from a defined, default status.
*RST
# Switch off the automatic display update to increase programming
# speed.
:DISPlay OFF
# Set the instrument to pattern mode and select NRZ data output format
# for output 1 and output2.
:DIGital:PATTern ON
:DIGital:SIGNa1l:FORMat NRZ
:DIGital:SIGNal2:FORMat NRZ
# Define a pattern length of 8 bit and program for channel 1 a pattern
# with clock division by 2, for channel 2 a pattern with clock division
# by 8.
:TRIGger:COUNt 8
:DIGital:PATTern:PRESet1 2,8
:DIGital:PATTern:PRESet2 8,8