Measurement
37
Transmission output
hold (current hold)
setting for
transmission output 2
Specify the transmission output value when transmission output
2 is in the hold mode.
HoLd: Holds and outputs the latest value
PrES: Outputs the preset value
HoLd,
PrES
Transmission output 2
hold output value
(preset) setting
Holds the preset value.
Note:
This is displayed when “Transmission output hold (current hold)
setting for transmission output 2” is set to PrES.
10.0 to
110.0
(
C)
Transmission output 2
burnout hold setting
Specify the output during burnout of transmission output 2.
out.4: Outputs 3.8 mA at the time of burnout
out.20: Outputs 21 mA at the time of burnout
non: No burnout setting
Note:
Transmission burnout appears only in the measurement mode.
non, out.4,
out.20
Cleaning operation setting
Reference
Refer to “ Electrochemical cleaning capability ” (page 77).
Specifies whether
cleaning is performed
when the power is
turned ON
Specify whether or not to perform cleaning when the power is
turned ON.
no: Cleaning is not performed
yES: Cleaning is performed
yES, no
Specifies whether
cleaning is performed
when the cleaning
flow switch is reset
Specify whether or not to perform cleaning when a flow error is
reset.
no: Cleaning is not performed
yES: Cleaning is performed
yES, no
Setting for
Electrochemical
cleaning
Electro-chemical cleaning will be activated or not.
no: Cleaning is stopped
yES: Cleaning is activated intermittently
yES, no
Specifies a cleaning
frequency
Specify a cleaning frequency in days and hours.
1, 2, 4, 6, 8,
12 (h), 1 to
7 (days)
Specifies cleaning
time setting
Specify cleaning time in seconds.
5 to 600 (s)
Specifies HOLD
output time after the
end of cleaning
Specify hold retention time in minutes after cleaning is finished.
1 to 30
(min)
RS-485 address setting
RS-485 address
setting
Specify the RS-485 communication address.
0 to 99
Display
Item
Description
Setting
range